3.10.4 Non-ionising imaging
Describe how ultrasound forms images, distinguish A-scans from B-scans, calculate the depth of a boundary, and evaluate the technique.
1/6Ultrasound images are built from echoes at tissue boundaries
- Ultrasound images are built from echoes at tissue boundaries
- A piezoelectric transducer sends pulses and detects echoes
- Coupling gel lets ultrasound enter the body
- A-scans show echo depth and B-scans build a picture
- Echo time and ultrasound speed give the depth of a boundary
- Ultrasound is safe and flexible but has clear limits
- Check your recall
- Practice: 10 questions
Part 1 of 6: Ultrasound images are built from echoes at tissue boundaries
Shout across a valley and you hear an echo. Ultrasound imaging uses the same idea with sound too high for people to hear. Sound with a frequency above 20 kHz is called ultrasound. Medical scanners use frequencies of a few megahertz.
A scanner sends short pulses of ultrasound into the body. At a boundary between two different tissues, part of each pulse is reflected. The rest carries on deeper. The reflected part is the echo, and the scanner detects it. The time each echo takes to return shows how far away the boundary is.
Key term
ultrasound
Ultrasound is sound with a frequency above the upper limit of human hearing, about 20 kHz.
Part 2 of 6: A piezoelectric transducer sends pulses and detects echoes
Now consider how the pulses are made. A transducer is the probe that sends and receives the ultrasound. It contains a crystal of piezoelectric material.
An alternating potential difference across the crystal makes it vibrate, so it emits ultrasound. When an echo reaches the crystal, it vibrates and produces a small potential difference. The same crystal therefore transmits and receives.
The transducer sends a very short pulse, then listens. Long gaps between pulses let echoes return before the next pulse starts.
Key terms · tap one for its definition
transducer
A transducer is a device that converts energy from one form to another; an ultrasound transducer converts between electrical signals and ultrasound.
piezoelectric material
A piezoelectric material is one that changes shape when a potential difference is applied and produces a potential difference when deformed.

Quick check
One question on the part you just read
Easy
Show answer and explanation
Answer: B · Worked explanation
Part 3 of 6: Coupling gel lets ultrasound enter the body
A pulse reflects wherever the material changes. How much reflects depends on acoustic impedance. For a material it is , where is the density and the speed of ultrasound in the material.
The greater the difference in impedance between two materials, the larger the fraction of intensity reflected. Air and soft tissue differ hugely. Almost all the ultrasound would therefore reflect at the skin.
A coupling gel is spread between the transducer and the skin. It removes the layer of air, and its impedance is close to that of skin. More ultrasound therefore enters the body.
Key terms · tap one for its definition
acoustic impedance
The acoustic impedance of a material is , the product of its density and the speed of sound in it, with unit .
coupling gel
A coupling gel is a gel with impedance close to that of skin, applied to remove air between the transducer and the body.
Part 4 of 6: A-scans show echo depth and B-scans build a picture
Now consider how the echoes are displayed. An A-scan uses one transducer in a fixed position. It plots echo amplitude against time along a single line through the body.
Each peak on the trace is an echo from a boundary. Later peaks come from deeper boundaries. An A-scan is useful for measuring distances, such as the length of the eye.
A B-scan combines many A-scans from different positions or directions. The echoes are shown as spots whose brightness represents echo strength. Together the spots form a two-dimensional image of a slice through the body.
Key terms · tap one for its definition
A-scan
An A-scan is a display of echo amplitude against time for ultrasound sent along a single line into the body.
B-scan
A B-scan is a two-dimensional ultrasound image built from many A-scans, with echo strength shown as brightness.

Quick check
One question on the part you just read
Easy
Show answer and explanation
Answer: B · Worked explanation
Part 5 of 6: Echo time and ultrasound speed give the depth of a boundary
The pulse travels to a boundary and back, so the recorded time covers twice the depth. Let be the time between the pulse and its echo. Let be the speed of ultrasound in the tissue. The depth is then .
In soft tissue, is about . Times are often given in microseconds, so convert to seconds first. A longer delay means a deeper boundary.
Worked example
Finding the depth of a boundary
- Depth is speed multiplied by time, divided by two.
- Write the equation: .
- Substitute: .
- Evaluate: .
Answer: The boundary is , or 3.9 cm, deep.
Part 6 of 6: Ultrasound is safe and flexible but has clear limits
An ultrasound wave is non-ionising. It cannot remove electrons from an atom to make an ion, so it is regarded as safe at imaging intensities.
Advantages and limitations of ultrasound imaging
- Advantages
- Non-ionising at imaging intensities
- Limitations
- Bone and air block the view of structures behind
- Advantages
- Images appear in real time
- Limitations
- Detail is limited by the wavelength used
- Advantages
- Equipment is cheap and portable
- Limitations
- Higher frequencies penetrate less deeply
- Advantages
- Shows soft tissue boundaries well
- Limitations
- Gel and good contact are needed
A higher frequency has a shorter wavelength, so it resolves finer detail. But it is also absorbed more strongly, so it cannot reach deep structures. Operators therefore choose a frequency to suit the depth.
Bone and air reflect strongly. Structures behind them are hard to see. The operator must also keep the probe in good contact with the skin.
Key terms · tap one for its definition
non-ionising
A non-ionising wave does not carry enough energy to remove electrons from atoms, so it does not form ions in the tissue it crosses.
ion
An ion is an atom or group of atoms with a net charge from losing or gaining electrons.
Check your recall[16 marks]
6 written questions. Answer each one in your head or on paper, then open the mark scheme.
- 1.Describe[3 marks]
Describe how a transducer produces ultrasound pulses and detects echoes.
Show mark schemeHide mark scheme
Model answer
A transducer contains a piezoelectric crystal. An alternating potential difference makes the crystal vibrate and emit a pulse of ultrasound. When an echo reaches the crystal, it vibrates and produces a potential difference that is detected.
- 1 mark for stating that the transducer contains a piezoelectric crystal
- 1 mark for stating that an alternating potential difference makes the crystal vibrate to emit pulses
- 1 mark for stating that returning echoes make the crystal produce a potential difference
- 2.Explain[2 marks]
Explain why a coupling gel is used between the transducer and the skin.
Show mark schemeHide mark scheme
Model answer
Air and tissue have very different acoustic impedances, so almost all the ultrasound would reflect at the skin. The gel removes the air and has an impedance close to skin, so more ultrasound enters the body.
- 1 mark for stating that the gel removes air, which would reflect almost all the ultrasound
- 1 mark for stating that the gel’s impedance is close to that of skin, so more ultrasound enters
- 3.Compare[3 marks]
Compare an A-scan with a B-scan.
Show mark schemeHide mark scheme
Model answer
An A-scan plots echo amplitude against time along one line. A B-scan combines many A-scans into a two-dimensional image. In a B-scan, echo strength is shown as brightness instead of peak height.
- 1 mark for stating that an A-scan plots amplitude against time along a single line
- 1 mark for stating that a B-scan forms a two-dimensional image from many A-scans
- 1 mark for stating that a B-scan shows echo strength as brightness
- 4.Calculate[3 marks]
Calculate the depth of a boundary when an echo returns after the pulse. Use for the speed of ultrasound.
Show mark schemeHide mark scheme
Model answer
.
- 1 mark for using with the factor of two
- 1 mark for converting 64 µs to
- 1 mark for obtaining
- 5.Give[2 marks]
Give two advantages of ultrasound imaging.
Show mark schemeHide mark scheme
Model answer
It is non-ionising, so it does not form ions in tissue. It gives images in real time. The equipment is cheap and portable.
- 1 mark for giving one valid advantage, such as being non-ionising
- 1 mark for giving a second valid advantage, such as real-time images or portable equipment
- 6.Suggest[3 marks]
Suggest why higher-frequency ultrasound suits structures near the skin and lower-frequency ultrasound suits deep structures.
Show mark schemeHide mark scheme
Model answer
Higher frequency has a shorter wavelength, so it resolves finer detail. It is absorbed more strongly, so it does not reach deep structures. Lower frequency penetrates further, so it suits deep structures.
- 1 mark for stating that higher frequency has a shorter wavelength and resolves finer detail
- 1 mark for stating that higher frequency is absorbed more strongly and penetrates less
- 1 mark for stating that lower frequency penetrates further, so it suits deep structures

Practice this topic
10 quick-fire multiple-choice questions, easy to hard. Tap an answer to see if you're right and the worked explanation. (The written questions with mark schemes are under “Check your recall” above.)
0 right0 of 10 answered
Includes the 2 quick checks in the note above.
EasyState[1 mark]
Show answer and explanation
Answer: C · Worked explanation
MediumCalculate[2 marks]
Show answer and explanation
Answer: B · Worked explanation
MediumCompare[2 marks]
Show answer and explanation
Answer: C · Worked explanation
MediumIdentify[2 marks]
Show answer and explanation
Answer: B · Worked explanation
MediumCalculate[2 marks]
Show answer and explanation
Answer: D · Worked explanation
HardPredict[2 marks]
Show answer and explanation
Answer: A · Worked explanation
HardCalculate[2 marks]
Show answer and explanation
Answer: D · Worked explanation
HardExplain[2 marks]
Show answer and explanation
Answer: D · Worked explanation