3.10.4 Non-ionising imaging

Describe how ultrasound forms images, distinguish A-scans from B-scans, calculate the depth of a boundary, and evaluate the technique.

  • 5 min read
  • 6 parts
  • 10 questions
  • Draft
Start a Quick 5 on this topic→Jump to the 10 questions
1/6Ultrasound images are built from echoes at tissue boundaries
  1. 1Ultrasound images are built from echoes at tissue boundaries
  2. 2A piezoelectric transducer sends pulses and detects echoes
  3. 3Coupling gel lets ultrasound enter the body
  4. 4A-scans show echo depth and B-scans build a picture
  5. 5Echo time and ultrasound speed give the depth of a boundary
  6. 6Ultrasound is safe and flexible but has clear limits
  7. Check your recall
  8. Practice: 10 questions
1

Part 1 of 6: Ultrasound images are built from echoes at tissue boundaries

Shout across a valley and you hear an echo. Ultrasound imaging uses the same idea with sound too high for people to hear. Sound with a frequency above 20 kHz is called ultrasound. Medical scanners use frequencies of a few megahertz.

A scanner sends short pulses of ultrasound into the body. At a boundary between two different tissues, part of each pulse is reflected. The rest carries on deeper. The reflected part is the echo, and the scanner detects it. The time each echo takes to return shows how far away the boundary is.

Key term

ultrasound

Ultrasound is sound with a frequency above the upper limit of human hearing, about 20 kHz.

Part 1 done

2

Part 2 of 6: A piezoelectric transducer sends pulses and detects echoes

Now consider how the pulses are made. A transducer is the probe that sends and receives the ultrasound. It contains a crystal of piezoelectric material.

An alternating potential difference across the crystal makes it vibrate, so it emits ultrasound. When an echo reaches the crystal, it vibrates and produces a small potential difference. The same crystal therefore transmits and receives.

The transducer sends a very short pulse, then listens. Long gaps between pulses let echoes return before the next pulse starts.

Key terms · tap one for its definition

transducer

A transducer is a device that converts energy from one form to another; an ultrasound transducer converts between electrical signals and ultrasound.

piezoelectric material

A piezoelectric material is one that changes shape when a potential difference is applied and produces a potential difference when deformed.

Part 2 done

Quick check

One question on the part you just read

Easy

Identify the frequency above which sound is classed as ultrasound.
Show answer and explanation

Answer: B · Worked explanation

Ultrasound is sound above the upper limit of human hearing, about 20 kHz. The value 20 Hz is the lower limit of hearing, so sounds near it are very low pitched. Medical scanners use a few megahertz, which is far above this limit and is also ultrasound.
3

Part 3 of 6: Coupling gel lets ultrasound enter the body

A pulse reflects wherever the material changes. How much reflects depends on acoustic impedance. For a material it is Z=ρcZ=\rho cZ=ρc, where ρ\rhoρ is the density and ccc the speed of ultrasound in the material.

The greater the difference in impedance between two materials, the larger the fraction of intensity reflected. Air and soft tissue differ hugely. Almost all the ultrasound would therefore reflect at the skin.

A coupling gel is spread between the transducer and the skin. It removes the layer of air, and its impedance is close to that of skin. More ultrasound therefore enters the body.

Exam tip

When you explain the gel, say that it removes air and reduces the impedance difference at the skin.

Key terms · tap one for its definition

acoustic impedance

The acoustic impedance of a material is Z=ρcZ=\rho cZ=ρc, the product of its density and the speed of sound in it, with unit kg m−2 s−1\mathrm{kg\,m^{-2}\,s^{-1}}kgm−2s−1.

coupling gel

A coupling gel is a gel with impedance close to that of skin, applied to remove air between the transducer and the body.

Part 3 done

4

Part 4 of 6: A-scans show echo depth and B-scans build a picture

Now consider how the echoes are displayed. An A-scan uses one transducer in a fixed position. It plots echo amplitude against time along a single line through the body.

Each peak on the trace is an echo from a boundary. Later peaks come from deeper boundaries. An A-scan is useful for measuring distances, such as the length of the eye.

Figure 1. The transmitted pulse is followed by echoes from boundaries. Deeper boundaries give echoes at later times.

A B-scan combines many A-scans from different positions or directions. The echoes are shown as spots whose brightness represents echo strength. Together the spots form a two-dimensional image of a slice through the body.

Key terms · tap one for its definition

A-scan

An A-scan is a display of echo amplitude against time for ultrasound sent along a single line into the body.

B-scan

A B-scan is a two-dimensional ultrasound image built from many A-scans, with echo strength shown as brightness.

Part 4 done

Quick check

One question on the part you just read

Easy

Name the type of material in a transducer that vibrates when a potential difference is applied across it.
Show answer and explanation

Answer: B · Worked explanation

A piezoelectric material changes shape when a potential difference is applied, so it vibrates when the potential difference alternates. It also produces a potential difference when an echo deforms it. A conducting gel is used between transducer and skin, but it does not vibrate.
5

Part 5 of 6: Echo time and ultrasound speed give the depth of a boundary

The pulse travels to a boundary and back, so the recorded time covers twice the depth. Let ttt be the time between the pulse and its echo. Let ccc be the speed of ultrasound in the tissue. The depth is then d=ct2d=\frac{ct}{2}d=2ct​.

In soft tissue, ccc is about 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1. Times are often given in microseconds, so convert to seconds first. A longer delay means a deeper boundary.

Worked example

Finding the depth of a boundary

An echo returns 52 μs52\,\mathrm{\mu s}52μs after a pulse. The speed of ultrasound in the tissue is 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1. Calculate the depth of the boundary.
  1. 1
    Depth is speed multiplied by time, divided by two.
  2. 2
    Write the equation: d=ct2d=\frac{ct}{2}d=2ct​.
  3. 3
    Substitute: d=1.5×103 m s−1×52×10−6 s2d=\frac{1.5\times10^{3}\,\mathrm{m\,s^{-1}}\times52\times10^{-6}\,\mathrm{s}}{2}d=21.5×103ms−1×52×10−6s​.
  4. 4
    Evaluate: d=0.078 m2=3.9×10−2 md=\frac{0.078\,\mathrm{m}}{2}=3.9\times10^{-2}\,\mathrm{m}d=20.078m​=3.9×10−2m.

Answer: The boundary is 3.9×10−2 m3.9\times10^{-2}\,\mathrm{m}3.9×10−2m, or 3.9 cm, deep.

Common mistake

Do not use the full echo time. That gives twice the true depth, because the pulse travels there and back.

Part 5 done

6

Part 6 of 6: Ultrasound is safe and flexible but has clear limits

An ultrasound wave is non-ionising. It cannot remove electrons from an atom to make an ion, so it is regarded as safe at imaging intensities.

Advantages and limitations of ultrasound imaging

AdvantagesLimitations
Non-ionising at imaging intensitiesBone and air block the view of structures behind
Images appear in real timeDetail is limited by the wavelength used
Equipment is cheap and portableHigher frequencies penetrate less deeply
Shows soft tissue boundaries wellGel and good contact are needed
Advantages
Non-ionising at imaging intensities
Limitations
Bone and air block the view of structures behind
Advantages
Images appear in real time
Limitations
Detail is limited by the wavelength used
Advantages
Equipment is cheap and portable
Limitations
Higher frequencies penetrate less deeply
Advantages
Shows soft tissue boundaries well
Limitations
Gel and good contact are needed

A higher frequency has a shorter wavelength, so it resolves finer detail. But it is also absorbed more strongly, so it cannot reach deep structures. Operators therefore choose a frequency to suit the depth.

Bone and air reflect strongly. Structures behind them are hard to see. The operator must also keep the probe in good contact with the skin.

Key terms · tap one for its definition

non-ionising

A non-ionising wave does not carry enough energy to remove electrons from atoms, so it does not form ions in the tissue it crosses.

ion

An ion is an atom or group of atoms with a net charge from losing or gaining electrons.

Note read

Check your recall[16 marks]

6 written questions. Answer each one in your head or on paper, then open the mark scheme.

  1. 1.Describe[3 marks]

    Describe how a transducer produces ultrasound pulses and detects echoes.

    Show mark schemeHide mark scheme

    Model answer

    A transducer contains a piezoelectric crystal. An alternating potential difference makes the crystal vibrate and emit a pulse of ultrasound. When an echo reaches the crystal, it vibrates and produces a potential difference that is detected.

    • 1 mark for stating that the transducer contains a piezoelectric crystal
    • 1 mark for stating that an alternating potential difference makes the crystal vibrate to emit pulses
    • 1 mark for stating that returning echoes make the crystal produce a potential difference
  2. 2.Explain[2 marks]

    Explain why a coupling gel is used between the transducer and the skin.

    Show mark schemeHide mark scheme

    Model answer

    Air and tissue have very different acoustic impedances, so almost all the ultrasound would reflect at the skin. The gel removes the air and has an impedance close to skin, so more ultrasound enters the body.

    • 1 mark for stating that the gel removes air, which would reflect almost all the ultrasound
    • 1 mark for stating that the gel’s impedance is close to that of skin, so more ultrasound enters
  3. 3.Compare[3 marks]

    Compare an A-scan with a B-scan.

    Show mark schemeHide mark scheme

    Model answer

    An A-scan plots echo amplitude against time along one line. A B-scan combines many A-scans into a two-dimensional image. In a B-scan, echo strength is shown as brightness instead of peak height.

    • 1 mark for stating that an A-scan plots amplitude against time along a single line
    • 1 mark for stating that a B-scan forms a two-dimensional image from many A-scans
    • 1 mark for stating that a B-scan shows echo strength as brightness
  4. 4.Calculate[3 marks]

    Calculate the depth of a boundary when an echo returns 64 μs64\,\mathrm{\mu s}64μs after the pulse. Use 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1 for the speed of ultrasound.

    Show mark schemeHide mark scheme

    Model answer

    d=ct2=1.5×103×64×10−62=4.8×10−2 md=\frac{ct}{2}=\frac{1.5\times10^{3}\times64\times10^{-6}}{2}=4.8\times10^{-2}\,\mathrm{m}d=2ct​=21.5×103×64×10−6​=4.8×10−2m.

    • 1 mark for using d=ct2d=\frac{ct}{2}d=2ct​ with the factor of two
    • 1 mark for converting 64 µs to 64×10−6 s64\times10^{-6}\,\mathrm{s}64×10−6s
    • 1 mark for obtaining 4.8×10−2 m4.8\times10^{-2}\,\mathrm{m}4.8×10−2m
  5. 5.Give[2 marks]

    Give two advantages of ultrasound imaging.

    Show mark schemeHide mark scheme

    Model answer

    It is non-ionising, so it does not form ions in tissue. It gives images in real time. The equipment is cheap and portable.

    • 1 mark for giving one valid advantage, such as being non-ionising
    • 1 mark for giving a second valid advantage, such as real-time images or portable equipment
  6. 6.Suggest[3 marks]

    Suggest why higher-frequency ultrasound suits structures near the skin and lower-frequency ultrasound suits deep structures.

    Show mark schemeHide mark scheme

    Model answer

    Higher frequency has a shorter wavelength, so it resolves finer detail. It is absorbed more strongly, so it does not reach deep structures. Lower frequency penetrates further, so it suits deep structures.

    • 1 mark for stating that higher frequency has a shorter wavelength and resolves finer detail
    • 1 mark for stating that higher frequency is absorbed more strongly and penetrates less
    • 1 mark for stating that lower frequency penetrates further, so it suits deep structures

Practice this topic

10 quick-fire multiple-choice questions, easy to hard. Tap an answer to see if you're right and the worked explanation. (The written questions with mark schemes are under “Check your recall” above.)

0 right0 of 10 answered

Includes the 2 quick checks in the note above.

3.

Easy·State[1 mark]

State why a coupling gel is placed between the transducer and the skin.
Show answer and explanation

Answer: C · Worked explanation

Air and skin have very different acoustic impedances, so air would reflect almost all the ultrasound. The gel removes the air and has an impedance close to skin. More ultrasound then enters the body. The gel does not change the frequency, which is set by the transducer.
4.

Medium·Calculate[2 marks]

An echo returns 40 μs40\,\mathrm{\mu s}40μs after a pulse. The speed of ultrasound in the tissue is 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1. Calculate the depth of the boundary.
Show answer and explanation

Answer: B · Worked explanation

d=ct2=1.5×103×40×10−62=3.0×10−2 m=3.0 cmd=\frac{ct}{2}=\frac{1.5\times10^{3}\times40\times10^{-6}}{2}=3.0\times10^{-2}\,\mathrm{m}=3.0\,\mathrm{cm}d=2ct​=21.5×103×40×10−6​=3.0×10−2m=3.0cm. The value 6.0 cm comes from forgetting to halve the time. That gives the total distance travelled by the pulse, there and back, instead of the depth of the boundary below the skin.
5.

Medium·Compare[2 marks]

Which statement correctly compares an A-scan with a B-scan?
Show answer and explanation

Answer: C · Worked explanation

An A-scan is a trace of echo amplitude against time along one line. A B-scan combines many A-scans and shows echo strength as brightness in a two-dimensional image. Reversing the two descriptions is a common error. Neither technique uses ionising radiation.
6.

Medium·Identify[2 marks]

Identify what the brightness of a spot represents in a B-scan.
Show answer and explanation

Answer: B · Worked explanation

In a B-scan the position of each spot comes from the echo time and the direction of the pulse. Its brightness shows how strong the echo was. The time taken is tempting, but the time only fixes the spot’s position, so it cannot also set the brightness.
7.

Medium·Calculate[2 marks]

A boundary is 6.0 cm6.0\,\mathrm{cm}6.0cm below the skin. The speed of ultrasound in the tissue is 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1. Calculate the time for the echo to return.
Show answer and explanation

Answer: D · Worked explanation

The pulse travels there and back, a distance of 2×0.060 m2\times0.060\,\mathrm{m}2×0.060m. Then t=2dc=0.121.5×103=8.0×10−5 s=80 μst=\frac{2d}{c}=\frac{0.12}{1.5\times10^{3}}=8.0\times10^{-5}\,\mathrm{s}=80\,\mathrm{\mu s}t=c2d​=1.5×1030.12​=8.0×10−5s=80μs. The value 40 µs comes from using only the one-way distance, which ignores the return journey of the echo back to the transducer.
8.

Hard·Predict[2 marks]

A scanner changes from ultrasound of frequency 3 MHz to ultrasound of frequency 6 MHz. Predict the effect on the image.
Show answer and explanation

Answer: A · Worked explanation

At the same speed, a higher frequency means a shorter wavelength, so finer detail can be resolved. Higher frequencies are also absorbed more strongly, so penetration falls. The belief that neither changes ignores wavelength, which depends on frequency even when the speed is constant.
9.

Hard·Calculate[2 marks]

In soft tissue where the speed of ultrasound is 1.5×103 m s−11.5\times10^{3}\,\mathrm{m\,s^{-1}}1.5×103ms−1, echoes from two boundaries return 30 μs30\,\mathrm{\mu s}30μs and 70 μs70\,\mathrm{\mu s}70μs after the pulse. Calculate the distance between the boundaries.
Show answer and explanation

Answer: D · Worked explanation

The extra time is 70−30=40 μs70-30=40\,\mathrm{\mu s}70−30=40μs. Then d=ct2=1.5×103×40×10−62=3.0×10−2 m=3.0 cmd=\frac{ct}{2}=\frac{1.5\times10^{3}\times40\times10^{-6}}{2}=3.0\times10^{-2}\,\mathrm{m}=3.0\,\mathrm{cm}d=2ct​=21.5×103×40×10−6​=3.0×10−2m=3.0cm. The value 5.3 cm is the depth of the deeper boundary alone, measured from the skin, so it is not the distance between the two boundaries.
10.

Hard·Explain[2 marks]

Explain why ultrasound gives poor images of structures behind air-filled lungs.
Show answer and explanation

Answer: D · Worked explanation

Air and soft tissue have very different acoustic impedances. Almost all the ultrasound is therefore reflected at the boundary, and little reaches structures beyond. The claim that air has the same impedance as tissue is wrong. If it were true, little would be reflected there and no echo would form.

Keep going

Next topic3.10.5 X-ray imagingPrevious topic3.10.3 Biological measurement

More in Section 3.10 · Option: Medical physics

  1. 3.10.1Physics of the eyeNotes · 10 questions
  2. 3.10.2Physics of the earNotes · 10 questions
  3. 3.10.3Biological measurementNotes · 10 questions
  4. 3.10.4Non-ionising imagingNotes · 10 questions
  5. 3.10.5X-ray imagingNotes · 10 questions
  6. 3.10.6Radionuclide imaging and therapyNotes · 10 questions

All AQA A-level Physics topics →

Draft: written by Scribe and checked automatically, not yet by a subject reviewer.

Practice questions are original, written by Scribe and checked automatically for correctness.

Topic codes follow the official numbering. Not endorsed by AQA or any exam board.

On this page

0/6 read

  1. 1Ultrasound images are built from echoes at tissue boundaries
  2. 2A piezoelectric transducer sends pulses and detects echoes
  3. 3Coupling gel lets ultrasound enter the body
  4. 4A-scans show echo depth and B-scans build a picture
  5. 5Echo time and ultrasound speed give the depth of a boundary
  6. 6Ultrasound is safe and flexible but has clear limits