Scribe logoScribeNotes & practice
Sign inStart free
  1. Notes
  2. ›AQA A-level Physics
  3. ›3.10 Option: Medical physics
  4. ›3.10.5 X-ray imaging

AQA A-level Physics · Section 3.10 · Option: Medical physics

3.10.5 X-ray imaging

Describe how X-rays are produced and attenuated, compare soft and hard X-rays, calculate attenuation, and explain the use of contrast media.

  • 4 min read
  • 5 parts
  • 10 questions
  • Draft
Start a Quick 5 on this topic→Jump to the 10 questions
1/5Fast electrons hitting a metal target produce X-rays
  1. 1Fast electrons hitting a metal target produce X-rays
  2. 2Tissues attenuate X-rays by absorbing and scattering photons
  3. 3Soft X-rays are absorbed easily, while hard X-rays penetrate
  4. 4Intensity falls exponentially as X-rays pass through material
  5. 5Contrast media make soft tissues visible on X-ray images
  6. Check your recall
  7. Practice: 10 questions
1

Part 1 of 5: Fast electrons hitting a metal target produce X-rays

A current heats a filament at one end of a sealed, evacuated tube. The hot metal releases electrons by thermionic emission.

A high potential difference, typically tens of kilovolts, accelerates the electrons towards a metal target at the other end. When the electrons stop in the target, some of their kinetic energy becomes X-ray photons.

Only a small fraction of the energy becomes X-rays. Most heats the target. It is therefore made from a metal with a high melting point and is cooled, often by rotating it.

Key terms · tap one for its definition

thermionic emission

Thermionic emission is the release of electrons from the surface of a metal when it is heated to a high temperature.

target

The target is the metal anode of an X-ray tube that the electron beam strikes and from which X-rays are emitted.

Part 1 done

2

Part 2 of 5: Tissues attenuate X-rays by absorbing and scattering photons

As X-rays cross the body, some photons are absorbed or scattered. The loss of intensity this causes is attenuation. Different tissues attenuate by different amounts, so the emerging beam carries an image.

  • The photoelectric effect: a photon is absorbed and ejects an electron from an atom. It is strong in materials of high atomic number.
  • Compton scattering: a photon gives some energy to an electron and moves off in a new direction.
  • Pair production: a photon above 1.02 MeV1.02\,\mathrm{MeV}1.02MeV becomes an electron and a positron. Diagnostic X-rays have too little energy.

Common mistake

Attenuation includes scattering as well as absorption. Do not describe it as absorption only.

Bone has a higher atomic number than soft tissue, and it is denser. It absorbs more X-rays and appears clearly on the image.

Key terms · tap one for its definition

attenuation

Attenuation is the reduction in intensity of a beam of X-rays as it passes through matter, caused by absorption and scattering.

photoelectric effect

The photoelectric effect is the process in which a photon is absorbed by an atom and its energy ejects an electron.

Compton scattering

Compton scattering is the process in which a photon transfers part of its energy to an electron and continues with lower energy.

Part 2 done

Quick check

One question on the part you just read

Easy

Name the process by which a heated filament in an X-ray tube releases electrons.
Show answer and explanation

Answer: B · Worked explanation

Thermionic emission is the release of electrons from the surface of a hot metal. The photoelectric effect is tempting because it also releases electrons, but it needs photons to be absorbed by atoms. In the tube it is heat, not photons, that frees the electrons.
3

Part 3 of 5: Soft X-rays are absorbed easily, while hard X-rays penetrate

Now consider how photon energy affects the beam. A soft X-ray photon has low energy and a long wavelength. A hard X-ray photon has high energy and a short wavelength.

Soft and hard X-rays compared

FeatureSoft X-raysHard X-rays
Photon energyLowerHigher
WavelengthLongerShorter
PenetrationAbsorbed in the first tissuesPass through the body
Effect on imageAdd dose, no imageForm the image
Feature
Photon energy
Soft X-rays
Lower
Hard X-rays
Higher
Feature
Wavelength
Soft X-rays
Longer
Hard X-rays
Shorter
Feature
Penetration
Soft X-rays
Absorbed in the first tissues
Hard X-rays
Pass through the body
Feature
Effect on image
Soft X-rays
Add dose, no image
Hard X-rays
Form the image

Soft X-rays are absorbed in the body and add to the patient’s dose without helping the image. A thin aluminium filter in the beam removes most of them. A higher tube voltage makes the beam harder. Very hard X-rays pass through almost everything, so tissues differ less in absorption and contrast falls.

Go deeperWhy a higher voltage gives harder X-rays

An electron accelerated through a potential difference VVV gains kinetic energy eVeVeV, where eee is the elementary charge. The most energetic photon forms when an electron gives up all this energy at once, so hfmax=eVhf_\mathrm{max}=eVhfmax​=eV.

This gives a shortest wavelength of λmin=hceV\lambda_\mathrm{min}=\frac{hc}{eV}λmin​=eVhc​. Raising VVV makes this wavelength shorter, so the beam becomes harder.

Key terms · tap one for its definition

soft X-ray

A soft X-ray is a low-energy X-ray photon that tissue absorbs easily, so it penetrates the body poorly.

hard X-ray

A hard X-ray is a high-energy X-ray photon that penetrates tissue easily and passes through the body.

Part 3 done

4

Part 4 of 5: Intensity falls exponentially as X-rays pass through material

Now consider how much of a beam gets through a given thickness. Each equal thickness of a uniform material removes the same fraction of the intensity. This is an exponential decrease: I=I0e−μxI=I_0e^{-\mu x}I=I0​e−μx.

Here I0I_0I0​ is the intensity entering the material and III is the intensity after thickness xxx. The linear attenuation coefficient μ\muμ depends on the material and the photon energy. Its unit is m−1\mathrm{m^{-1}}m−1.

The half-value thickness x1/2x_{1/2}x1/2​ halves the intensity. Setting I=I02I=\frac{I_0}{2}I=2I0​​ gives x1/2=ln⁡2μx_{1/2}=\frac{\ln 2}{\mu}x1/2​=μln2​.

Figure 1. The intensity falls by the same fraction for each equal thickness. The half-value thickness is marked on the curve.

Worked example

Finding the transmitted intensity

A material has μ=0.35 cm−1\mu=0.35\,\mathrm{cm^{-1}}μ=0.35cm−1. An X-ray beam crosses 4.0 cm4.0\,\mathrm{cm}4.0cm of it. Calculate the percentage of the intensity transmitted.
  1. 1
    The transmitted fraction is II0=e−μx\frac{I}{I_0}=e^{-\mu x}I0​I​=e−μx.
  2. 2
    Find the exponent: μx=0.35 cm−1×4.0 cm=1.4\mu x=0.35\,\mathrm{cm^{-1}}\times4.0\,\mathrm{cm}=1.4μx=0.35cm−1×4.0cm=1.4.
  3. 3
    Evaluate: II0=e−1.4=0.247\frac{I}{I_0}=e^{-1.4}=0.247I0​I​=e−1.4=0.247.
  4. 4
    Convert to a percentage: 0.247×100%≈25%0.247\times100\%\approx25\%0.247×100%≈25%.

Answer: About 25% of the intensity is transmitted.

Exam tip

Use matching units. If μ\muμ is in cm−1\mathrm{cm^{-1}}cm−1, the thickness must be in cm.

Key terms · tap one for its definition

linear attenuation coefficient

The linear attenuation coefficient, μ\muμ, is the constant in I=I0e−μxI=I_0e^{-\mu x}I=I0​e−μx, measured in m−1\mathrm{m^{-1}}m−1, and is larger for more strongly attenuating material.

half-value thickness

The half-value thickness, x1/2x_{1/2}x1/2​, is the thickness of a material that reduces the X-ray intensity to half its initial value.

Part 4 done

Quick check

One question on the part you just read

Easy

State how the intensity of an X-ray beam changes as it passes through increasing thicknesses of a uniform material.
Show answer and explanation

Answer: A · Worked explanation

Each equal thickness removes the same fraction of the intensity, which gives an exponential decrease, I=I0e−μxI=I_0e^{-\mu x}I=I0​e−μx. A linear decrease would remove the same amount of intensity in each thickness, so the beam would reach zero at a fixed depth. The beam instead keeps falling by the same fraction.
5

Part 5 of 5: Contrast media make soft tissues visible on X-ray images

Soft tissues attenuate X-rays by similar amounts, so an organ such as the stomach shows little contrast. A contrast medium fixes this. It contains atoms of high atomic number, and the patient swallows it or is injected with it.

Atoms of high atomic number absorb X-rays strongly, mainly by the photoelectric effect. The filled organ therefore stands out clearly against its surroundings.

A barium meal contains barium sulfate and shows the digestive tract. Compounds of iodine are injected to show blood vessels.

Common mistake

A contrast medium does not emit X-rays. It shows up because it absorbs more of the beam.

Key term

contrast medium

A contrast medium is a substance of high atomic number introduced into the body to increase X-ray absorption in a chosen organ.

Note read

Check your recall[15 marks]

5 written questions. Answer each one in your head or on paper, then open the mark scheme.

  1. 1.Describe[4 marks]

    Describe how X-rays are produced in an X-ray tube.

    Show mark schemeHide mark scheme

    Model answer

    A heated filament releases electrons by thermionic emission. A high potential difference accelerates them across a vacuum. They strike a metal target. Some of their kinetic energy becomes X-ray photons, and most becomes heat.

    • 1 mark for stating that a heated filament releases electrons by thermionic emission
    • 1 mark for stating that a high potential difference accelerates the electrons
    • 1 mark for stating that the electrons strike a metal target
    • 1 mark for stating that some kinetic energy becomes X-ray photons and most becomes heat
  2. 2.Compare[3 marks]

    Compare soft X-rays and hard X-rays.

    Show mark schemeHide mark scheme

    Model answer

    Soft X-rays have lower photon energy and longer wavelength than hard X-rays. Soft X-rays are absorbed easily by tissue. Hard X-rays penetrate the body and form the image.

    • 1 mark for stating that soft X-rays have lower photon energy and longer wavelength
    • 1 mark for stating that soft X-rays are absorbed easily in tissue
    • 1 mark for stating that hard X-rays penetrate the body and form the image
  3. 3.Calculate[3 marks]

    Calculate the fraction of the intensity transmitted through 3.0 cm3.0\,\mathrm{cm}3.0cm of a material with μ=0.45 cm−1\mu=0.45\,\mathrm{cm^{-1}}μ=0.45cm−1.

    Show mark schemeHide mark scheme

    Model answer

    II0=e−μx=e−0.45×3.0=e−1.35=0.26\frac{I}{I_0}=e^{-\mu x}=e^{-0.45\times3.0}=e^{-1.35}=0.26I0​I​=e−μx=e−0.45×3.0=e−1.35=0.26.

    • 1 mark for using I=I0e−μxI=I_0e^{-\mu x}I=I0​e−μx
    • 1 mark for calculating the exponent μx=1.35\mu x=1.35μx=1.35
    • 1 mark for obtaining a fraction of 0.26
  4. 4.Explain[3 marks]

    Explain why a barium meal allows the stomach to be seen on an X-ray image.

    Show mark schemeHide mark scheme

    Model answer

    Soft tissues attenuate X-rays by similar amounts, so the stomach shows little contrast. Barium has a high atomic number and absorbs X-rays strongly. The filled stomach therefore stands out against the surrounding tissue.

    • 1 mark for stating that soft tissues attenuate X-rays by similar amounts
    • 1 mark for stating that barium has a high atomic number and absorbs X-rays strongly
    • 1 mark for stating that the filled stomach stands out against its surroundings
  5. 5.State[2 marks]

    State what is meant by the half-value thickness of a material and give its relationship with μ\muμ.

    Show mark schemeHide mark scheme

    Model answer

    It is the thickness that reduces the intensity to half its initial value. x1/2=ln⁡2μx_{1/2}=\frac{\ln 2}{\mu}x1/2​=μln2​.

    • 1 mark for stating that it is the thickness that halves the intensity
    • 1 mark for giving x1/2=ln⁡2μx_{1/2}=\frac{\ln 2}{\mu}x1/2​=μln2​

Practice this topic

10 quick-fire multiple-choice questions, easy to hard. Tap an answer to see if you're right and the worked explanation. (The written questions with mark schemes are under “Check your recall” above.)

0 right0 of 10 answered

Includes the 2 quick checks in the note above.

3.

Easy·Identify[1 mark]

Identify the substance used in a barium meal to show the digestive tract.
Show answer and explanation

Answer: C · Worked explanation

A barium meal contains barium sulfate. Barium has a high atomic number, so it absorbs X-rays strongly and outlines the digestive tract. Aluminium is tempting because it appears in X-ray work, but it is used as a filter in the beam to remove soft X-rays, not to show organs.
4.

Medium·Calculate[2 marks]

A beam crosses 5.0 cm5.0\,\mathrm{cm}5.0cm of a material with μ=0.20 cm−1\mu=0.20\,\mathrm{cm^{-1}}μ=0.20cm−1. Calculate the percentage of the intensity transmitted.
Show answer and explanation

Answer: D · Worked explanation

The exponent is μx=0.20×5.0=1.0\mu x=0.20\times5.0=1.0μx=0.20×5.0=1.0. Then II0=e−1.0=0.37\frac{I}{I_0}=e^{-1.0}=0.37I0​I​=e−1.0=0.37, so 37% is transmitted. The value 63% is the fraction removed from the beam, not the fraction transmitted, because 1−0.37=0.631-0.37=0.631−0.37=0.63. The exponent of 1.0 shows that the transmitted fraction is e−1.0e^{-1.0}e−1.0, which is well below one half.
5.

Medium·Calculate[2 marks]

A material has μ=0.50 cm−1\mu=0.50\,\mathrm{cm^{-1}}μ=0.50cm−1. Calculate its half-value thickness.
Show answer and explanation

Answer: C · Worked explanation

x1/2=ln⁡2μ=0.6930.50 cm−1=1.4 cmx_{1/2}=\frac{\ln 2}{\mu}=\frac{0.693}{0.50\,\mathrm{cm^{-1}}}=1.4\,\mathrm{cm}x1/2​=μln2​=0.50cm−10.693​=1.4cm. The value 2.0 cm comes from using 1μ\frac{1}{\mu}μ1​, which is the thickness that reduces the intensity to 1e\frac{1}{e}e1​ of its value, not to a half. The half-value thickness needs the factor ln⁡2\ln 2ln2, which is about 0.693, in the numerator.
6.

Medium·Explain[2 marks]

Explain why bone appears clearly against soft tissue on an X-ray image.
Show answer and explanation

Answer: B · Worked explanation

Bone is denser than soft tissue and contains atoms of higher atomic number, so it attenuates more X-rays, mainly by the photoelectric effect. The difference in attenuation gives contrast. The idea that bone emits X-rays is wrong, because attenuation removes photons from the beam and emits none.
7.

Medium·Explain[2 marks]

Explain why an aluminium filter is placed in the beam from an X-ray tube.
Show answer and explanation

Answer: C · Worked explanation

Soft X-rays have low photon energy and are absorbed in the first tissues, adding to the patient’s dose without reaching the detector. The filter removes most of them, so the dose falls but the image is unchanged. Removing the hard X-rays would be wrong, because they form the image.
8.

Hard·Predict[2 marks]

The potential difference across an X-ray tube is doubled. Predict the effect on the shortest wavelength of the X-rays produced.
Show answer and explanation

Answer: B · Worked explanation

The shortest wavelength is λmin=hceV\lambda_\mathrm{min}=\frac{hc}{eV}λmin​=eVhc​, so it is inversely proportional to VVV. Doubling VVV halves λmin\lambda_\mathrm{min}λmin​. The belief that it is doubled treats the wavelength as proportional to VVV, but a more energetic photon has a shorter wavelength, not a longer one.
9.

Hard·Calculate[2 marks]

A 3.0 cm3.0\,\mathrm{cm}3.0cm thickness of a material transmits 40% of the intensity of an X-ray beam. Calculate the linear attenuation coefficient.
Show answer and explanation

Answer: D · Worked explanation

From II0=e−μx\frac{I}{I_0}=e^{-\mu x}I0​I​=e−μx, 0.40=e−3.0μ0.40=e^{-3.0\mu}0.40=e−3.0μ. Taking logarithms, 3.0μ=ln⁡(10.40)=0.9163.0\mu=\ln\left(\frac{1}{0.40}\right)=0.9163.0μ=ln(0.401​)=0.916, so μ=0.31 cm−1\mu=0.31\,\mathrm{cm^{-1}}μ=0.31cm−1. The value 0.92 is the product μx\mu xμx itself. It forgets to divide by the thickness, so it is a pure number and cannot have units of inverse centimetres.
10.

Hard·Identify[2 marks]

Identify the correct statement about a contrast medium used in X-ray imaging.
Show answer and explanation

Answer: C · Worked explanation

A contrast medium contains atoms of high atomic number, which absorb X-rays strongly. The organ it fills therefore attenuates more than the tissue around it. A low atomic number would give less attenuation and so less contrast. Contrast media absorb X-rays and do not emit them.

Keep going

Next topic3.10.6 Radionuclide imaging and therapyPrevious topic3.10.4 Non-ionising imaging

More in Section 3.10 · Option: Medical physics

  1. 3.10.1Physics of the eyeNotes · 10 questions
  2. 3.10.2Physics of the earNotes · 10 questions
  3. 3.10.3Biological measurementNotes · 10 questions
  4. 3.10.4Non-ionising imagingNotes · 10 questions
  5. 3.10.5X-ray imagingNotes · 10 questions
  6. 3.10.6Radionuclide imaging and therapyNotes · 10 questions

All AQA A-level Physics topics →

Draft: written by Scribe and checked automatically, not yet by a subject reviewer.

Practice questions are original, written by Scribe and checked automatically for correctness.

Topic codes follow the official numbering. Not endorsed by AQA or any exam board.

On this page

0/5 read

  1. 1Fast electrons hitting a metal target produce X-rays
  2. 2Tissues attenuate X-rays by absorbing and scattering photons
  3. 3Soft X-rays are absorbed easily, while hard X-rays penetrate
  4. 4Intensity falls exponentially as X-rays pass through material
  5. 5Contrast media make soft tissues visible on X-ray images
  6. Check your recall
  7. Practice: 10 questions
All coursesScribeHow it worksPricingPrivacyTerms

Notes and questions are written by Scribe in our own words. Topic codes follow the official course numbering so you can find your place. Not produced or endorsed by the IB, College Board, AQA, OCR, Pearson Edexcel or any exam board.

© 2026 Scribe

Intensity against thickness

Transmitted intensity (% of initial)012345678020406080100Thickness of material (cm)
  • Transmitted intensity
  • Half-value thickness